Cookies del sito web Omega Utilizziamo i cookies nel nostro sito web. Questi cookies sono necessari per consentire al sito web di funzionare correttamente. Se si continua senza modificare la propria configurazione del browser, si acconsente a ricevere i cookies dal nostro sito web. Per maggiori chiarimenti sull'utilizzo dei cookies, cliccare qui
Calibrate Thermocouple Instruments to 0.1 and 0.01, °F and °C Types J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), (T-DIN) and P (Platinel): -13.000 to 80.000 mV
Calibrate RTD Instruments to 0.1 and 0.01, °F and °C Pt 10, 50, 100, 200, 500 and 1000O (3850) Pt 100O (alpha = 3902, 3916, 3926) Cu 10 and 50O and Ni 120O
Calibrate with Confidence Resolution to 0.01, °F and °C; Accurate to 0.2° Cold Junction Sensor Accurate to 0.05°C NIST Traceable Cold Junction Sensor in Isothermal Block
Troubleshoot RTD Sensors and Instruments
Automatic Detection of 2-, 3- or 4-Wire Connections Indicates When Individual Wire Connections are Open Verify Proper Instrument Operation by Measuring the Excitation Current
Fast Easy Calibrations Instantly Select 3 Outputs with the EZ-CHECK Switch Easily Find Trip Points with 2 Speed EZ-DIAL Knob and Automatic Stepping
Verify Resistance, Cold Junction Temperature and Output Secondary Display of Resistance for RTDs and Compensated millivolt Setting, and Cold Junction Temperature for Thermocouple
Use in the Plant, Field or Shop Includes Rubber Boot and mV Test Leads Low Profile Switches Resist Accidental Damage Brass Screws for Thermocouple Wires Plus Miniature Thermocouple Socket
Perform Heat Treating Surveys and Accuracy Tests Meets or Exceeds the Requirement of AMS 2750 as Both a Secondary Standard Instrument and as a Field Test Instrument
THERMOCOUPLE CONNECTIONS
Simulating or reading
thermocouples requires
the use of thermocouple
or extension grade
thermocouple wire.
Plug thermocouple wires into
the miniature thermocouple
jack or place bare
thermocouple wires
onto the brass block under
the screws.
The CL543B has two banana
jacks (1+ and 2-) mounted in
the top end of the housing.
These are not temperature
compensated and are to be
used only for millivolt signals.
RTD CONNECTIONS
Simulating or reading RTDs
uses copper wire. Plug 2-, 3- or
4-wires into the corresponding
jacks on the calibrator. For RTD
source the CL543B simulates the
(+) RTD from jacks 1 and 4 and
the (-) RTD from jacks 2 and 3.
When reading an RTD sensor
the CL543B uses patented
circuitry to automatically detects
if 2-, 3- or 4-wires are connected.
This is helpful to troubleshoot
sensor connections.
Specifications Operating Temperature Range:
-20 to 60°C Storage Temperature Range:
-30 to 60°C Temperature Effect: ≤±25 ppm/°C of
range Cold Junction Sensor: ≤±0.01% °/°
from ambient temperature (°C or °F) Relative Humidity Range:
10% ≤RH ≤90% (0 to 35°C), non-condensing
10% ≤RH ≤70% (35 to 60°C), non-condensing Normal Mode Rejection: 50/60 Hz, 50 dB Common Mode Rejection: 50/60 Hz, 120 dB Dimensions: 143 L x 76 W x 41 mm H Weight: 0.34 kg with boot
and batteries Batteries: Four “AA” alkaline 1.5V (LR6) (included), Battery Life: 50 hours, low battery
indication with nominal 1 hour of
operation left Over-Voltage Protection: Up to
60 vrms (rated for 30 seconds),
protection against misconnection Alerts: Red LED displays
OVERLOAD or out of range conditions Display: High contrast graphic liquid
crystal display, LED backlighting for
use in low lit areas Voltage Source Ranges and Resolution:
-13.000 to 80.000 mV and -13.0000
to 80.0000 mV Accuracy: ±(0.008% of setting +
0.006 mV) Source Current: ≥10 mA Output Impedance: <0.3Ω RMS Noise: ≤±0.0005 mV from
0.1 to 10 Hz Short Circuit Duration: Infinite
Voltage Read Ranges and Resolution:
-13.000 to 80.000 mV and -13.0000
to 80.0000 mV Accuracy: ±(0.008% of reading +
0.006 mV) Input resistance: ≥10 MΩ
Thermocouple Source Accuracy: ±(0.008% of setting +
0.006 mV) Cold Junction Compensation:
±0.05°C (±0.09°F); thermistor
traceable to NIST for 11 years Output Impedance: <0.3Ω Source Current: >10 mA (drives
80 mV into 10Ω) RMS Noise: ≤±0.0005 mV from
0.1 to 10 Hz Thermocouple Read Accuracy: ±(0.008% of reading +
0.006 mV) Cold Junction Compensation:
±0.09°F (±0.05°C); thermistor
traceable to NIST for 11 years Input Impedance: >10 MΩ Open Thermocouple Threshold:
10,000Ω nominal Pulse: <10 microamp pulse for
300 milliseconds RTD and O Source 3- and 4-Wire Accuracy: From 1 to 10.2 mA External
Excitation Current: ±(0.015% of setting + 0.05Ω) Below 1 mA of External Excitation Current:
Add (0.025 mV/mA excitation
current) 2-Wire Accuracy: Add 0.1Ω to 3-wire
and 4-wire accuracy Resistance Ranges: 400Ω Range: 0.00 to 401.00 and
0.000 to 401.000 4000Ω Range: 0.0 to 4010.0 and
0.00 to 4010.00 RMS Noise: 400Ω Range: ≤±0.005Ω from
0.1 to 10 Hz 4000Ω Range: ≤±0.05Ω from
0.1 to 10 Hz Allowable Excitation Current Range: 400Ω Range: 10.2 mA max; steady
or pulsed/intermittent 4000Ω Range: 1 mA max; steady
or pulsed/intermittent Pulsed Excitation Current
Compatibility: DC to 0.01 second
pulse width, 0.025 mV, mA excitation
current RTD and Ohms Read Resistance Ranges: 400Ω Range: 0.00 to 401.00 and
0.000 to 401.000 4000Ω Range: 0.0 to 4010.0 and
0.00 to 4010.00 Accuracy: ±(0.015% of reading +
0.05Ω) Excitation Current: 0.9 mA to 401Ω,
0.4 mA to 4010Ω (nominal)
† Tutti gli importi indicati in EUR Nota: Comes complete with NIST certificate, 4 “AA” alkaline batteries, certificate of calibration card, rubber boot, deluxe hands free carrying case,
mA/V test leads, 1 red and 1 black lead with banana plugs and alligator clips, RTD wire kit, 2 red and 2 black leads with banana plugs and
spade lugs, and operator’s manual.