Vedi tutti i modelli qui sotto
-
As an alternative see the PX02 series or the “Pressure Transducer Builder”
-
Stainless Steel Diaphragm for High Media Compatibility
-
Rugged Stainless Steel Case Protects Components in Industrial Environments
-
Supplied with 36" Color Coded, Shielded Leads for Easy Connections and High Noise Immunity
-
Uses a Standard 10 Vdc Regulated Power Supply for Maximum Versatility
-
Compact Design for Use in Restricted Areas
-
1 /4 NPT Female Process Thread for Easy Installations on Piping Systems
Trasduttori di pressione - Visualizza i prodotti correlati
Descrizione
This item is being discontinued.
SPECIFICATIONS
Excitation: 10 Vdc (15 Vdc max.)
Output: 30 mV @ 10 Vdc
Sensitivity: 3 mV/V nominal
Input Impedance: 360 ohm
Output Resistance: 350 ohm
Insulation Resistance: 5 MOhm @ 75 Vdc
PERFORMANCE
Accuracy: ±0.3% (includes linearity, hysteresis repeatability)
Zero Balance: ±3%
Operating Temperature Range: -65 to 250°F (-54 to 121°C)
Compensated Temperature Range: 60 to 160°F (16 to 71°C)
Thermal Zero Effect: ±0.01% full scale/°F
Thermal Sensitivity Effect: ±0.02% full scale/°F
Proof Pressure:150% of Range
Burst Pressure: 300% of Range
CONSTRUCTION
Body Material: 17-4 PH Stainless Steel
Diaphragm Material: 17-4 PH Stainless Steel
Pressure Port: 1 /4 -18 NPT female
Electrical Connection: 3 ft (0.9 m) 4 conductor shielded cable 3 /8 -18 NPT male conduit fitting with 11 /16 hex
Weight: 8 oz (227 g)