Dissolved Oxygen - The Fundamentals
Introduction
Dissolved
oxygen (DO) is the term commonly used in liquid analytical work
for the measurement of the amount of oxygen dissolved in a unit
volume of water. It is an important indicator of the degree of
usefulness of a sample of water for a specific application. The
requirements of a given application determine the level of DO
that can be tolerated.
In
a water quality application, for example, where we want to maintain
a fresh water stream fit for recreational purposes such as swimming
and fishing and as a source of potable water, we must keep the
DO content high. If the DO level falls too low, the fish will
suffocate and conditions will become favorable for the growth
of harmful bacteria.
In
sewage treatment, solids are allowed to settle in large basins
to which are added solutions rich in bacteria to speed the decomposition
of the solids. There is an optimum DO level for this process and
the level is maintained by mechanically aerating the "activated
sludge" as the bacteria impregnated content of the basins is called.
If the DO level falls too low, the bacteria die and the decomposition
ceases; if the DO level is excessive, more power is used than
necessary for aeration and the process becomes unnecessarily costly.
Another important application of DO is the control of the quality
of boiler make up water. In this case, since the presence of oxygen
in the water will enhance corrosion and cause the build up of
boiler scale that inhibits heat transfer, it is very desirable
to hold the DO concentration to a minimum.
Theory
The
amount of oxygen that a given volume of water can hold is a function
of:
1. The pressure the atmospheric oxygen is exerting at the air-
water interface.
2. The temperature of the water.
3. The amount of other substances dissolved in the water.
Effect
of Partial Pressure of Oxygen on Dissolved Oxygen
A volume of water in contact with air will absorb air and hence
oxygen until the pressure the absorbed oxygen exerts at the air-water
interface equals the pressure exerted at the same interface by
the oxygen in the air. At this point, the water is said to be
saturated with oxygen. The amount of oxygen actually absorbed
is quite small being of the order of about five or ten parts of
oxygen to one million parts of water.
Effect
of Temperature on Dissolved Oxygen
As everyone who has ever watched a pot boil knows, bubbles form
on the side and bottom of the pot. The number and size of the
bubbles increase with temperature. These are bubbles of air that
have been dissolved in the water. Figure 1-A represents a beaker
of oxygen saturated water at room temperature. The partial pressures
of oxygen above and below the water surface are equal. Figure
1-B shows what happens when we begin to heat the system. In essence,
we are putting energy into the system. The oxygen molecules which
have a low solubility become readily energized by the infusion
of energy into the system. The more sluggish water molecules step
up their molecular activity at a slower pace. As a result, more
oxygen molecules break through the air-water interface to the
space above the water surface than water molecules do, leaving
fewer oxygen molecules dissolved in the water. When the water
boils as is shown in Figure 1-C, all of the oxygen molecules have
been driven out of the water and now the water molecules are coming
out of the beaker so fast that they form a layer of water vapor
immediately above the water surface. This layer effectively isolates
the water from atmospheric oxygen and the DO content of the water
is zero.
We
can think of a volume of water as if it were a homogeneous medium
with a number of holes interspersed throughout the volume. The
pressure of the air immediately above the water surface will cause
these holes to be filled with air. If now a substance such as
salt is dissolved in the water, the dissolved salt will occupy
some of the holes used by the oxygen molecules as they constantly
shift from air to water and water to air in an equilibrium situation.
The amount of oxygen the water can hold at given temperature has
now decreased, but the partial pressure of the dissolved oxygen
remaining in the solution must still equal the partial pressure
of the atmospheric oxygen above the water surface. This situation
is graphically portrayed in Figure 2 where curve A represents
fresh water and curve B represents salt water.
Measurement
Techniques
Basically there are two general techniques for measuring DO. Each
employs an electrode system wherein the dissolved oxygen reacts
at the cathode producing a measurable electrochemical effect.
The effect may be galvanic, polarographic or potentiometric.
One
technique uses a Clark-type cell which is merely an electrode
system separated from the sample stream by a semi-permeable membrane.
This membrane permits the oxygen dissolved in the sample to pass
through it to the electrode system while preventing liquids and
ionic species from doing so. The cathode is a hydrogen electrode
and carries a negative applied potential with respect to the anode.
Electrolyte surrounds the electrode pair and is contained by the
membrane. In the absence of a reactant, the cathode becomes polarized
with hydrogen and resistance to current flow becomes infinite.
When a reactant, such as oxygen that has passed through the membrane
is present, the cathode is depolarized and electrons are consumed.
The anode of the electrode pair must react with the product of
the depolarization reaction with a corresponding release of electrons.
As a result, the electrode pair permits current to flow in direct
proportion to the amount of oxygen or reactant entering the system;
hence, the magnitude of the current gives us a direct measure
of the amount of oxygen entering the system.
Membrane
probes readily lend themselves to conditions of high interfacial
turbulence. In the case of the thallium probe, a high degree of
turbulence may decrease the life of the probe because high turbulence
will sweep away the thallium ions, thus causing electrode depletion.
Although
dissolved organic materials are not known to interface with the
output from dissolved oxygen probes, inorganic salts are a factor
in the performance of the probes. As we saw in Figure 2, we must
apply a correction factor to the output of probes with membranes
when used in salt solutions. The thallium probe requires the presence
of salts in concentrations which provide a minimum conductivity
of approximately 200 micromhos.
The
two principle gaseous interferers of membrane probe measurements
are chlorine and hydrogen sulfide. Sulfur compounds such as hydrogen
sulfide, sulfur dioxide and mercaptans cause erroneous outputs
from the thallium probe. Halogens do not interfere with the thallium
probe.Figure 3 - Velocity, Agitation, and/or Turbulence Effects
of Interfacial Dynamics on Probe Output.Figure 2 - Typical Curves
Concentration of Dissolved Oxygen at Saturation
At
low dissolved oxygen concentrations, pH variations below pH 5
and above pH 9 interfere with the performance of the thallium
probe. This interference amounts to about ±0.5 mg/L DO per pH
unit. The performance of membraned probes is not affected by pH
changes. A serious limitation of the thallium probe is the fact
that thallium is quite toxic and you must exercise care in using
it. Membrane probes do not have this drawback. Since all of the
oxygen that passes through the membrane reacts and since the amount
of oxygen that passes through the membrane is a function of the
partial pressure of the oxygen in solution, this technique actually
measures the partial pressure of the oxygen in solution. It does
not measure the actual concentration of the oxygen in the solution.
For this reason, we must correct the readings of DO concentration
given by this technique when some substance, for example salt,
is dissolved in the water. As we saw above, the dissolved salt
will reduce the number of holes available for carrying oxygen
and hence reduce the actual concentration of oxygen without changing
its partial pressure in the solution. If the electrode materials
are selected so that the difference in potential is -0.5 volts
or greater at the cathode, an external potential is not required
and we have what is called a galvanic system. Some workers in
this field, instead of relying on their selection of electrode
materials to give them the required -0.5 volts difference of potential
at the cathode, use an external potential source to give them
the required potential difference. This system is known as a polarographic
system. In either case, since the partial pressure of dissolved
oxygen is a function of the temperature of the sample, we must
either hold the temperature of the sample constant or compensate
for varying sample temperature. Generally, the former is impractical
so the latter is the more popular approach. A suitably selected
thermistor or resistance thermometer in a properly designed electric
circuit does a fair job of temperature compensation.
The
second basic measuring technique uses an electrode system that
consists of a reference electrode and a thallium measuring electrode.
No semi-permeable membrane is used; the electrode system is immersed
directly into the sample. Oxygen concentration is determined by
measuring the voltage potential developed, in relation to the
reference electrode, when dissolved oxygen comes in contact with
the thallium electrode. At the surface of the electrode the thallous-ion
concentration is proportional to the dissolved oxygen. The voltage
potential developed by the cell is dependent upon the thallous-ion
concentration in this layer and varies as the dissolved oxygen
concentration changes. The cell output rises 59 millivolts for
each decade rise in oxygen concentration. This technique uses
a potentiometric system. The method measures directly the concentration
of oxygen in the sample. As in the first technique, temperature
compensation is a must and is achieved in about the same way.
In both techniques, interfacial dynamics at the probe-sample interface
are a factor in the probe response. A significant amount of interfacial
turbulence is necessary and for precision performance, turbulence
should be constant. This situation is portrayed in Figure 3. As
long as the operating point remains above the knee of the curve,
small changes in turbulence can be tolerated.
Curve
A 0 mg/L chlorides Curve B 5,000 mg/L chlorides Document #1234On-line
Publishing Service U.S.A. and Canada For Technical Assistance:
Reproduced
with permission of Royce Instruments.
|