« Tornare alla pagina intera
NUMERO VERDE 800 906 907 | INTERNAZIONALE +39 022 333 1521

Temperature Calibrator - Discontinued

CL123 CL125 and CL100MA

1 anno di garanzia CE RoHS
 
Temperature Calibrator - Discontinued
CL123 CL125 and CL100MA : Temperature Calibrator - Discontinued



   
 
Simulates Temperature or Process Signals
Handheld Temperature Calibrator Model CL123 for Simulation of mA, mV, V, Hz, or Thermocouple type K, J, E, T (in °C or °F) Signals
Handheld Thermocouple Calibrator Model CL125 for Simulation of Thermocouple type K, J, E, T, R, S, N, L, U, B, C (in °C, °F or K) Signals
Handheld Process Calibrator Model CL100MA for Simulation of 4-20 mA, 0-20 mA, 0-24 mA type Signals
Easy Push Button Programming

This product has been discontinued. For a possible replacement please see CL309A. Questions? Feel free to call one of our application engineers to help find the solution that is best for you.



The OMEGA™ CL100 Series are compact, easy-to-use handheld calibrators that provide a precision output signal. Models includes thermocouple calibrators, temperature calibrators, or process calibrators. Units do not read signals. Models are available that specialize in simulating thermocouples, voltage and output signals.

General Specifications
Battery: 9V Alkaline (included)
Displays: 4 and 5 digits
Operating Temperature:0 to 50°C (32 to 122°F)
Storage Temperature:-20 to 60°C (-4 to 140°F)
Operating/Storage Humidity:< 85% relative
Dimensions:88 x 168 x 26 mm (3.5 x 6.6 x 1")
Weight: 330 g (11.6 oz.)

CL123
Features
4-20 mA (1K. load, 24 V loop supply)
0-100.00 mV, 0-1.0000 V, 0-12.000 V
K,J,E,T Thermocouples (°C & °F)
Frequency 1-62500 Hz
0-20 mA, 0-24 mA Selectable
0-100% range setting in mA, mV, V, mode
Beeper Warning when Output is Open (mA)
Basic Accuracy: 0.025%
0-24 mA      ±0.025%   ±0.003 mA
0-100 mV    ±0.05%     ±0.03 mV
0-1 V           ±0.05%     ±0.0003 V
0-10 V         ±0.05%     ±0.003 V
Thermocouples:
K: -200 to 1370, ±0.9°C
J: -100 to 760, ±0.8°C
E: -100 to 700, ±0.7°C
T: -200 to 400, ±0.8°C
Frequency: 1 to 125 Hz, ±0.04 Hz, 126 to 62,500 Hz, ±0.01% ±0.04 Hz (604 Available Frequencies)
Thermocouple Connection:SMP (miniature) type on front
Process Connection:45 cm (18") red and black leads attached to side of unit

CL125
Features

11 Types of Thermocouples
°C, °F, K Temperature Scales
Internal/External Temperature Reference
Auto-Ramp and .T Function
Easy Keypad Programming
Basic Accuracy: 0.3°C/0.6°F
Resolution: 0.1°C/0.2°F (K,J,E,T,N,U,L)
0.2°C/0.4°F (R,S,B,C)
Range:
K: -200 to 1370°C       R: 0 to 1767°C
J: -200 to 1200°C       S: 0 to 1767°C
E: -200 to 1000°C      B: 600 to 1800°C
T: -200 to 400°C        N: -200 to 1300°C
C: 0 to 2300°C           U: -200 to 600°C
L: -200 to 900°C
Thermocouple Connection:SMP (miniature) type on front

CL100MA
Features
4-20 mA with 1 µA resolution
Auto-Ramp and Easy Step Functions
1-100% Input
0-20 mA, 0-24 mA Selectable
Easy Keypad Programming
RangeResolutionAccuracy
0-4 mA0.001 mA0.025% + 10 µA
4-20 mA0.001 mA0.025% + 5 µA
20-24 mA0.001 mA0.025% + 5 µA
Process Connection: 45 cm (18") red and black leads attached to side of unit



Ordinare (Specificare il codice prodotto)
expand list Quello altre persone hanno comprato: Quando vedi questa icona , fare cliccaci per espandere l'elenco dei prodotti che altre persone hanno comprato quando hanno acquistato questo modello.
     Codice prodotto     Disponibilitá   Prezzo   Descrizione   RoHS   
Fare clic per espandere / ridurre un elenco di ciò che gli altri hanno acquistato con il CL123  CL123     Consultare le vendite   Multifunction Handheld Temperature Calibrator and Process Calibrator Conforme a RoHS  
Fare clic per espandere / ridurre un elenco di ciò che gli altri hanno acquistato con il CL125  CL125     Consultare le vendite   Handheld, Temperature Calibrator, Thermocouple Calibrator Conformità RoHS indeterminata  
    CL100MA     Consultare le vendite   mA loop calibrator Conformità RoHS indeterminata  
 Accessories
Fare clic per espandere / ridurre un elenco di ciò che gli altri hanno acquistato con il CL123-ADAPTOR-115V  CL123-ADAPTOR-115V     Consultare le vendite   115 V power supply for CL123 and CL100MA Conformità RoHS indeterminata  
    CL123-ADAPTOR-230V     Consultare le vendite   230 V power supply for CL123 and CL100MA, European Plug Conforme a RoHS  


Nota:
Each unit supplied with rugged carrying case, complete operator’s manual, 9 V battery, external battery pack (without batteries) holds four AA batteries (models CL123 & CL100MA), type K thermocouple (models CL123, CL125), test leads and alligator clip (models CL123 & CL100MA). Units are for simulation (not reading) of temperature or process signals.

Esempio di ordine:
(1) CL125 Handheld, Temperature Calibrator, Thermocouple Calibrator, EUR 0,00

   
Parole chiavi:Multifunction Handheld Calibrators including Temperature Calibrator, Thermocouple Calibrator, and Process Calibrator, calibration labs, calibraion laboratories